Download

Download Manuals, Datasheets, Software and more:

DOWNLOAD TYPE
MODEL or KEYWORD

Feedback

Can I merge my TLA700 modules?

Question :

Can I merge my TLA700 modules?

Answer :

It depends whether or not the modules are compatible and only modules with 102 or more channels can merge. The following are the basic rules for merging modules:
  • TLA7Nx modules can be merged together, or with either TLA7Px or TLA7Qx modules.
  • TLA7Nx, TLA7Px and TLA7Qx modules cannot be merged with TLA7Lx and TLA7Mx modules.
  • TLA7Axx/TLA7NAx modules cannot be merged with TLA7Nx, TLA7Px, TLA7Qx, TLA7Lx, or TLA7Mx modules.
  • Any combination of TLA7Axx and TLA7NAx modules (up to five) can be merged.
  • Modules containing different memory depths can be merged but will default to the shallowest memory depth being used.
  • The maximum merged combinations are 5 TLA7Axx and TLA7NAx, 3 TLA7Nx/Px/Qx, or 2 TLA7Lx/Mx. The following graphic illustrates how to orient modules in a multiple module merge.
  • The number of channels in the master module must be equal to or greater than the number of channels in the module(s) with which it is merged. If there is a second slave module, the first slave must have greater than or equal to the number of channels in the second slave module. Slave 3 cannot have more channels than the master, slave 1, or slave 2. Slave 4 cannot have more channels than the master, slave 1, slave 2, or slave 3.
  • Modules cannot be merged between a mainframe and an expansion mainframe; merged modules must be in the same physical mainframe.
  • To merge a module to an established merged set, the established merged set must first be unmerged through software. Only unmerged modules can be used to create a merged configuration.

This FAQ Applies to:

Product Series: TLA7000

Product:

FAQ ID 57146

View all FAQs »