MSO4000+ DPO4000+ MSO3000+ DPO3000: For Measurements/Statistics the Mean & Std Dev Samples tab+ what does the number which is entered for Samples represent?

Question:

MSO4000+ DPO4000+ MSO3000+ DPO3000: For Measurements/Statistics the Mean & Std Dev Samples tab+ what does the number which is entered for Samples represent?

Answer:

The number you enters for Samples represents the number of waveforms acquired. If you enter the number 5 the scope will use the first 5 waveforms acquired. When 6 waveforms are acquired the scope will now uses waveforms 2-6 and does not use waveform number 1 which was the first acquisition. When the scope reaches 7 acquisitions it no longer uses the first or second waveform acquisitions and so forth.

This FAQ Applies to:

Product Series: MSO3000 / DPO3000, MSO3000 / DPO3000, MSO3000 / DPO3000, MSO3000 / DPO3000, MSO3000 / DPO3000, MSO3000 / DPO3000, MSO3000 / DPO3000, MSO3000 / DPO3000, MSO3000 / DPO3000, MSO3000 / DPO3000, MSO3000 / DPO3000

Product: DPO4054, DPO4034, DPO4032, DPO4104, MSO4032, MSO4034, MSO4054, MSO4104, DPO3012, DPO3014, DPO3032, DPO3034, DPO3052, DPO3054, MSO3012, MSO3014, MSO3032, MSO3034, MSO3054

FAQ ID : 60976