Download

Download Manuals, Datasheets, Software and more:

DOWNLOAD TYPE
MODEL or KEYWORD

Feedback

How do you run the self tests on the 2465?

Question :

How do you run the self tests on the 2465?

Answer :

Hold down both the Delta V and Delta T push buttons. Now, also press the Slope button. This will bring up a new screen. Release all buttons. Now push the Coupling Up button once to run a series of all tests once. Any failures will be displayed. Note that an "04 failure" simply means it is out of calibration.


This FAQ Applies to:

No product series

Product:

FAQ ID 58311

View all FAQs »