Model 2460 Simple TSP example of ten 5V pulses at 10 milliseconds each.

Question :

Model 2460 Simple TSP example of ten 5V pulses at 10 milliseconds each.

Answer :

Here is  a simple TSP (Test Script Processor) script to create ten 5V pulses of 0.01 second in duration.

(reset)
smu.source.level = 5
smu.measure.range = 100e-6
trigger.model.setblock(1,trigger.BLOCK_SOURCE_OUTPUT ,SMU.ON)
trigger.model.setblock(2,trigger.BLOCK_DELAY_CONSTANT,10e-3)
trigger.model.setblock(3,trigger.BLOCK_SOURCE_OUTPUT ,smu.OFF)
trigger.model.setblock(4,trigger.BLOCK_DELAY_CONSTANT ,10e-3)
trigger.model.setblock(5,trigger.BLOCK_BRANCH_COUNTER ,10,1)
trigger.model.initiate()